Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/25/2014 | 1505 | 11/12/2014 | 2359 | Incorrect |
Subject: | ENA/CPC/C1 - Instrument problem |
DataStreams: | enaaoscpcfC1.a1 |
Description: | Vacuum was lost very briefly to the instrument for an unknown reason. The instrument then showed evidence of flooding by an increase in the frequency of the butanol fill cycles, increase in nozzle pressure, and decrease in orifice pressure. |
Suggestions: | The data during this period should not be used |
Measurements: | enaaoscpcfC1.a1: more
|