Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2015 | 1506 | 06/17/2015 | 1321 | Incorrect |
Subject: | MAO/ACSM/AETH/CCN/CPC/HTDMA/NEPHELOMETER/PASS/PSAP/SMPS/SP2/UHSAS/S1 - Manifold blower |
DataStreams: | maoaossp2S1.00, maoaospsap3wS1.b1, maoaosnephwetS1.b1, maoaosuhsasS1.a1, maoaoscpcuS1.a1, maoaospass3wS1.a1, maoaosaethS1.00, maoaoshtdmaS1.a1, maoaoscpcfS1.a1, maoaossmpsS1.a1, maoaosacsmS1.a1, maoaospsap3wS1.a1, maoaosnephdryS1.b1, maoaossmpstrhS1.a1, maoaosccn100S1.a1, maoaosnephdryS1.a1, maoaosaethS1.a1 |
Description: | Manifold blower pulls air through the inlet to the instrument. Without the blower, the air source is indeterminant. |
Suggestions: | Do not use data. |
Measurements: | maoaosuhsasS1.a1: more
maoaospass3wS1.a1: more maoaossmpstrhS1.a1: more maoaossmpsS1.a1: more maoaosccn100S1.a1: more maoaospsap3wS1.b1: more maoaosaethS1.a1: more maoaoshtdmaS1.a1: more maoaosnephdryS1.b1: more maoaosnephdryS1.a1: more maoaosacsmS1.a1: more maoaosnephwetS1.b1: more maoaoscpcfS1.a1: more maoaossp2S1.00: more maoaospsap3wS1.a1: more maoaoscpcuS1.a1: more maoaosaethS1.00: more |