Subject: | MAO/PASS/S1 - Instrument problem |
DataStreams: | maoaospass3wS1.a1
|
Description: | Incorrect/Suspect Data due to noise
Noise was present at the start of the deployment due to unrecoverable issues occurring
during shipment. Period 1 (P1) is for this period until when the instrument was removed for
repair. P2 is the period when the spare PASS was installed. P3 is the period of spare
PASS data after a repair was attempted in the field. P4 is when the original PASS was
reinstalled after repair at the manufacturer. |
Suggestions: | All flagged data should be used with extreme caution. Intercomparisons with other
measurements are strongly suggested. Data is noisy data with low signals due to internal
issues with laser optics/stability. |