Subject:
SGP/CSAPR/I7 - PhiDp Decreasing with Range
DataStreams:sgpcsaprvertI7.00, sgpcsaprsurI7.00, sgpcsaprsecI7.00, sgpcsaprrhiI7.00
Description:
The differential phase is decreasing with range during this period, it should be 
increasing with range.  There are also periods of obviously bad PhiDp caused by a problem with the 
digital receiver software.
Suggestions: 
Flip the sign of the Phidp values for periods where it is decreasing with range.