Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/12/2015 | 0000 | 09/05/2015 | 1859 | Incorrect |
Subject: | MAO/GNDRAD/M1 - IRT bias |
DataStreams: | maogndrad60sM1.b1, maogndrad20sM1.a0 |
Description: | The surface temperatures measured by the IRT had a positive bias compared with the PIR. The IRT measured ~10K higher than the calibration blackbody. The IRT was replaced. |
Suggestions: | The data may be able to be corrected by deriving the regression against the PIR. |
Measurements: | maogndrad20sM1.a0: more
maogndrad60sM1.b1: more |