Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/09/2015 | 1029 | 10/22/2015 | 0000 | Incorrect |
Subject: | OLI/MET/M1 - Noisy (and missing) CMH data |
DataStreams: | olimetM1.b1 |
Description: | The chilled mirror hygrometer data became noisy after self-checks during this period. Data are missing between 10/16 and 10/21 for service and calibration checks. The instrument was returned to service following calibration checks and minor connection repairs. |
Suggestions: | Please use temp_mean, rh_mean, and dew_point_mean variables in this datastream as replacement data. |
Measurements: | olimetM1.b1: more
|