DQRID : D151214.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/11/2015100112/12/20150058Suspect
12/10/2015111712/11/20150636Suspect
12/12/2015201712/13/20152100Suspect
12/08/2015173912/08/20152346Suspect
12/13/2015204112/14/20150903Suspect
more
Subject:
ENA/UHSAS/C1 - High counts on three lowest channels
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the UHSAS three lowest channels. Probably a contamination (a fiber on the 
jet nozzle) inside the optical chamber. Rest of the channels should not be affected.
Suggestions: 
Do not use the data from the three lowest channels.
Measurements:enaaosuhsasC1.a1:
  • size_distribution
more