DQRID : D160209.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/18/2015014011/20/20152300Incorrect
more
Subject:
AWR/HTDMA/M1 - No particle counts
DataStreams:awraoshtdmaM1.a1
Description:
No particle counts after the instrument was turned on. Probably no butanol flow to the 
instrument or possibly a disconnected pump. No meaningful data saved.
Suggestions: 
Do not use the data from this time period.
Measurements:awraoshtdmaM1.a1:
  • aerosol_concentration
  • total_concentration
more