DQRID : D170131.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/15/2017103301/24/20171418Incorrect
more
Subject:
ASI/ACSM/M1 - Critical orifice clog
DataStreams:asiaosacsmM1.a1, asiaosacsmM1.b1, asiaosacsmM1.00, asiaosacsmM1.01
Description:
The critical orifice clog was placed on hold pending a site visit by the mentor. The 
quadrupole ionizer, lens system, and Prizma electronics head were replaced on 2017-01-20. The 
instrument was pumped down and tuned 2017-01-21. A lens alignment was performed 
2017-01-23 and the instrument calibrated 2017-01-24.
Suggestions: 
Data are unavailable or incorrect. Do no use data.
Measurements:asiaosacsmM1.b1:
  • total_organics
  • ammonium
  • chloride
  • org_mx_err
  • sulfate
  • org_mx
  • nitrate
  • amus
more
asiaosacsmM1.a1:
  • nitrate
  • ammonium
  • total_organics
  • inlet_closed
  • flow_rate
  • chloride
  • vaporizer
  • sulfate
  • emission_current
  • inlet_open
more
asiaosacsmM1.00:
  • Raw data stream - documentation not supported
more
asiaosacsmM1.01:
  • Raw data stream - documentation not supported
more