Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/15/2017 | 1033 | 01/24/2017 | 1418 | Incorrect |
Subject: | ASI/ACSM/M1 - Critical orifice clog |
DataStreams: | asiaosacsmM1.a1, asiaosacsmM1.b1, asiaosacsmM1.00, asiaosacsmM1.01 |
Description: | The critical orifice clog was placed on hold pending a site visit by the mentor. The quadrupole ionizer, lens system, and Prizma electronics head were replaced on 2017-01-20. The instrument was pumped down and tuned 2017-01-21. A lens alignment was performed 2017-01-23 and the instrument calibrated 2017-01-24. |
Suggestions: | Data are unavailable or incorrect. Do no use data. |
Measurements: | asiaosacsmM1.b1: more
asiaosacsmM1.a1: more asiaosacsmM1.00: more asiaosacsmM1.01: more |