Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/20/2017 | 0409 | 02/22/2017 | 1644 | Incorrect |
Subject: | SGP/MFRSR/E9 - Motor control board failure |
DataStreams: | sgpmfrsrE9.a0, sgpmfrsrauxE9.a0, sgpmfrsrlangleyE9.c1, sgpmfrsrE9.b1, sgpmfrsrE9.00, sgpmfrsrlangplotE9.c1 |
Description: | Regular site visit noted that the shadow band was not rotating. They first tried power cycling the instrument to no avail. The motor control board was ultimately replaced which corrected the issue. |
Suggestions: | |
Measurements: | sgpmfrsrE9.00: more
sgpmfrsrE9.b1: more sgpmfrsrE9.a0: more sgpmfrsrlangleyE9.c1: more sgpmfrsrlangplotE9.c1: more sgpmfrsrauxE9.a0: more |