DQRID : D170320.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/04/2003000002/06/20072359Suspect
more
Subject:
SGP/MET/E9 - Temperature occasionally failed 6-month field calibration checks
DataStreams:sgpmetE9.b1
Description:
The temperature/humidity sensor was not calibrated annually during this time. The 
temperature reading failed the 6-month field calibration check four times during this period.
Upon removal, the field calibration check reported the temperature sensor may have been 
reading high (< 1 deg C).

In general, the relative humidity may have been subject to drift during this time. The 
manufacturer characterizes sensor drift at <1% / year.
Suggestions: 
Please contact the mentor for specific information regarding the field calibration 

check results, if necessary.
Measurements:sgpmetE9.b1:
  • vapor_pressure_mean
  • rh_mean
  • temp_mean
more