Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/15/2016 | 0000 | 03/22/2017 | 1630 | Incorrect |
Subject: | SGP/EBBR/E11 - Reprocess: Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp5ebbrE11.b1, sgp15ebbrE11.b1, sgp30ebbrE11.b1 |
Description: | A wiring mistake caused the top and bottom temperature measurements to be incorrect (top and bottom relative humidity probe temperatures used to determine vapor pressure were correct). Thus the sensible and latent heat fluxes were not correctly calculated. |
Suggestions: | |
Measurements: | sgp15ebbrE11.b1: more
sgp5ebbrE11.b1: more sgp30ebbrE11.b1: more |