DQRID : D170404.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/03/20171609Open Ended1609Missing
03/29/2017162003/30/20171442Missing
04/01/2017183804/03/20171335Missing
03/31/2017150504/01/20171757Missing
more
Subject:
OLI/AOSMET/CPC/NEPHELOMETER/PSAP/TRACEGAS/AOSMET/M1 - All serial instruments in the AOS 
are down
DataStreams:oliaosmetM1.00
Description:
The Moxa IP to serial server failed.  There is no communication to any of the serial 
instruments in the AOS, so the data is missing.  Initially it was able to be rebooted and then 
worked for a while, but after a day it no longer started up.  A replacement is will be 
installed as soon as possible.
Suggestions: 
Measurements:oliaosmetM1.00:
  • Raw data stream - documentation not supported
more