Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/17/2017 | 1000 | 04/18/2017 | 2200 | Incorrect |
Subject: | OLI/UHSAS/M1 - Contamination in the optics chamber |
DataStreams: | oliaosuhsasM1.a1 |
Description: | Something stuck to the inlet jet nozzle causing high counts in the lower size bins. Other size bins also potentially affected. Some data missing as the instrument went into safe mode. |
Suggestions: | Do not use the data from this time period. |
Measurements: | oliaosuhsasM1.a1: more
|