Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/01/2017 | 0709 | 05/01/2017 | 1329 | Missing |
Subject: | ENA/ACSM/C1 - Filament burned out |
DataStreams: | enaaosacsmC1.01, enaaosacsmC1.00, enaaosacsmC1.b1, enaaosacsmC1.a1 |
Description: | Filament #1 burned out 2017-05-11 7:09. Filament #2 was turned on at 2017-05-11 13:29. Mass spectra were collected, but there were no ions generated so there was no signal. Only detector noise was collected. |
Suggestions: | |
Measurements: | enaaosacsmC1.b1: more
enaaosacsmC1.a1: more enaaosacsmC1.00: more enaaosacsmC1.01: more |