DQRID : D000210.1
Start DateStart TimeEnd DateEnd Time
06/11/1999000006/29/19992145
more
Subject:
SGP/EBBR/E26 - Temperature from Left T/RH Probe Incorrect
DataStreams:sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1
Description:
The temperature from the left T/RH probe went offscale frequently.  This
temperature is used in the calculation of the vapor pressure.  Therefore,
top and bottom Thum measurements, left Thum, top and bottom vapor pressure 
measurements, vapor pressure gradients, and the resulting latent and 
sensible heat fluxes and bowen ratio were frequently incorrect for the 
period stated.
Suggestions: 
Measurements:sgp5ebbrE26.b1:
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Top vapor pressure(vapor_pressure_top)
  • Top T/RH sensor temperature(temp_trh_top)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
more
sgp15ebbrE26.b1:
  • Left T/RH sensor temperature (resistance ratio)(temp_trh_left_rr)
more
sgp30ebbrE26.b1:
  • Top vapor pressure(vapor_pressure_top)
  • Bowen ratio(bowen_ratio)
  • Top T/RH sensor temperature(temp_trh_top)
  • Surface sensible heat flux(sensible_heat_flux)
  • Surface latent heat flux(latent_heat_flux)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
more