Subject: | TWP/MWR/C1 - Rain on window |
DataStreams: | twpmwrtipC1.a1, twpmwrlosC1.b1
|
Description: | The data is subject to contamination due to rain on the sensor window
and should be used with caution because the heater/blower assembly that
prevents water from accumulating on the sensor's teflon window was out
of commission. |
Suggestions: | |
Measurements: | twpmwrlosC1.b1: - Mean 31.4 GHz sky brightness temperature(tbsky31)
- Total water vapor along LOS path(vap)
- IR Brightness Temperature(ir_temp)
- Water on Teflon window (1=WET, 0=DRY)(wet_window)
- Averaged total liquid water along LOS path(liq)
- Mean 23.8 GHz sky brightness temperature(tbsky23)
more
twpmwrtipC1.a1: - 31.4 GHz sky signal(tipsky31)
- 23.8 GHz sky signal(tipsky23)
- Correlation coefficients for tip curve at 31.4 GHz(r31)
- 23.8 GHz goodness-of-fit coefficient(r23)
- 31.4 GHz blackbody(bb31)
- 31.4 GHz blac2body+noise injection signal(bbn31)
- 23.8 GHz Blackbody signal(bb23)
- 23.8 GHz blackbody+noise injection signal(bbn23)
- Water on Teflon window (1=WET, 0=DRY)(wet_window)
- IR Brightness Temperature(ir_temp)
- Tip configuration number(tipn)
more
|