Subject: | NSA/MWR/C1 - Reprocess: calibration change |
DataStreams: | nsamwrlosC1.b1, nsamwrlosC1.a1
|
Description: | The thicker (1/2-inch) insulation was installed on the RF deck and the reference
temperature was reduced from 50 C to 30 C to prevent thermal instability in winter. A change in
the mixer temperature affects the instrument's calibration. A new calibration was
automatically derived and applied after 1500 valid tip curves were acquired. During the period
after the temperature was reset and before the calibration was updated, the previous
calibration was used. |
Suggestions: | Reprocess the data with the new calibration. |
Measurements: | nsamwrlosC1.a1: - Mean 23.8 GHz sky brightness temperature(tbsky23)
- Mean 31.4 GHz sky brightness temperature(tbsky31)
- Total water vapor along LOS path(vap)
- Averaged total liquid water along LOS path(liq)
more
nsamwrlosC1.b1: - Averaged total liquid water along LOS path(liq)
- Total water vapor along LOS path(vap)
- Mean 23.8 GHz sky brightness temperature(tbsky23)
- Mean 31.4 GHz sky brightness temperature(tbsky31)
more
|