SGP/MWR/C1 - IRT failure
DataStreams:sgpmwrlosC1.b1, sgpmwrlosC1.a1
Description:
The IRT was damaged by internal moisture causing sky temperature
measurements that are negatively biased compared to those from
the AERI. IRT#0517 was returned to the manufacturer for repair
and was replaced with spare IRT#1254.
Suggestions: 
Measurements:sgpmwrlosC1.a1:
  • Sky Infra-Red Temperature(sky_ir_temp)
more
sgpmwrlosC1.b1:
  • Sky Infra-Red Temperature(sky_ir_temp)