DQRID : D020117.23
Start DateStart TimeEnd DateEnd Time
11/05/2001160011/05/20011901
11/11/2001143011/11/20011531
09/28/2001183009/28/20011831
12/22/2001110012/22/20011331
12/06/2001171512/06/20012331
10/16/2001191510/16/20012001
06/22/2001110006/22/20011331
07/21/2001173007/21/20011801
07/12/2001111507/12/20011531
12/22/2001080012/22/20011001
11/01/2001011511/01/20010131
07/04/2001113007/05/20010231
06/22/2001144506/22/20011501
07/12/2001050007/12/20010531
12/27/2001203012/28/20010031
01/08/2002161501/08/20022101
12/28/2001183012/28/20011931
12/28/2001234512/29/20010031
09/28/2001193009/29/20010001
07/18/2001023007/18/20010301
08/04/2001171508/04/20011831
01/07/2002093001/07/20021031
07/20/2001144507/20/20011931
07/14/2001023007/14/20011501
07/19/2001104507/19/20011431
07/19/2001020007/19/20010301
10/20/2001154510/20/20011631
07/18/2001081507/18/20010901
12/22/2001011512/22/20010401
01/11/2002153001/11/20021831
01/11/2002011501/11/20020201
07/19/2001054507/19/20010801
10/31/2001141510/31/20011731
06/26/2001084506/27/20010331
07/12/2001030007/12/20010401
11/11/2001101511/11/20011201
07/19/2001090007/19/20010931
06/27/2001054506/30/20011701
12/21/2001200012/21/20012031
12/17/2001134512/17/20011531
08/03/2001150008/04/20010101
10/16/2001231510/17/20010001
11/10/2001173011/10/20011901
11/02/2001153011/02/20011731
12/27/2001180012/27/20011901
01/05/2002173001/05/20022301
more
Subject:
SGP/EBBR/E2 - Reprocess: E2 EBBR Soil Temperature #2 Incorrect
DataStreams:sgp30ebbrE2.b1, sgp15ebbrE2.b1
Description:
Soil Temperature #2 was intermittently incorrect.
The cause is unknown.

The indicated measurements that are calculated using 
soil temperature are also incorrect.

Soil moisture #2 (30 minute only) was also affected 
because version 6 of the EBBR datalogger program 
uses a multiplication of soil temperature and soil 
moisture as part of the adjustment of soil moisture for
soil temperature.  This was discovered to be a poor 
technique soon after it was implemented, but E2 will
continue to use version 6 until a replacement EBBR
unit is available.  When soil temperature #2 is large
negative, soil moisture #2 is often twice as large
as it should be.

EBBR unit changeout dates are listed on the EBBR 
instrument web page.
Suggestions: 
Recalculate the average soil heat flow (ave_shf),
latent heat flux (e), and sensible heat flux (h) 
using only soil sets 1, 3, 4, and 5, as follows:

ave_shf = (g1 + g3 + g4 + g5)/4

e = -(q + ave_shf)/(1 + Bowen)

h = -(q + e + ave_shf)

where q is net radiation and Bowen is the Bowen ratio.
Measurements:sgp30ebbrE2.b1:
  • Soil moisture 2, gravimetric(soil_moisture_2)
  • Surface soil heat flux 2(surface_soil_heat_flux_2)
  • Change in energy storage 2, 0-5 cm soil layer(energy_storage_change_2)
  • Soil temperature 2(soil_temp_2)
  • Surface latent heat flux(latent_heat_flux)
  • Surface soil heat flux, average of fluxes 1-5(surface_soil_heat_flux_avg)
  • Surface sensible heat flux(sensible_heat_flux)
more
sgp15ebbrE2.b1:
  • Soil temperature 2 (resistance ratio)(soil_temp_2_rr)
more

Close this window