Subject: | SGP/SMOS/E1 - E1 SMOS RH% VALUES SUSPECT |
DataStreams: | sgp1smosE1.a1, sgp30smosE1.b1, sgp1smosE1.b1, sgp30smosE1.a1
|
Description: | RH% values approximately 10% high at values of 70% and higher. Sensor replaced. |
Suggestions: | |
Measurements: | sgp30smosE1.b1: - Standard Deviation of Relative Humidity(sd_rh)
- Vapor Pressure (kiloPascals)(vap_pres)
- Relative Humidity (hybrid-b grid)(rh)
- Standard deviation of vapor pressure(sd_vap_pres)
more
sgp30smosE1.a1: - Relative Humidity (hybrid-b grid)(rh)
- Standard Deviation of Relative Humidity(sd_rh)
- Standard deviation of vapor pressure(sd_vap_pres)
- Vapor Pressure (kiloPascals)(vap_pres)
more
sgp1smosE1.b1: - Relative Humidity (hybrid-b grid)(rh)
- Vapor Pressure (kiloPascals)(vap_pres)
more
sgp1smosE1.a1: - Vapor Pressure (kiloPascals)(vap_pres)
- Relative Humidity (hybrid-b grid)(rh)
more
|