DQRID : D020129.12
Start DateStart TimeEnd DateEnd Time
10/08/2001200010/12/20011731
more
Subject:
SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect
DataStreams:sgp30ebbrE13.b1, sgp15ebbrE13.b1
Description:
The left T/RH Probe Temperature was often incorrect, causing the other indicated
measurements to be incorrect also.
Suggestions: 
The incorrect measurements cannot easily be recalculated.
Measurements:sgp15ebbrE13.b1:
  • Left T/RH sensor temperature (resistance ratio)(temp_trh_left_rr)
more
sgp30ebbrE13.b1:
  • Bowen ratio(bowen_ratio)
  • Top T/RH sensor temperature(temp_trh_top)
  • Surface sensible heat flux(sensible_heat_flux)
  • Surface latent heat flux(latent_heat_flux)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
more

Close this window