Subject: | SGP/EBBR/E13 - Reprocess: E13 EBBR Soil Temperature #4 Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1
|
Description: | Soil Temperature #4 was often incorrect, typically too large negative,
during this period, causing other measurements calculated with
it to be incorrect also. This was caused by a bad multiplexor channel,
which was discovered by the vendor during recalibration of the unit
later. |
Suggestions: | The soil, sensible and latent heat fluxes can be recalculated using the
information from the four good soil sets and the equations below.
ave_shf = (g1 + g2 + g3 + g5)/4
e = -(ave_shf + q)/(1 + Bowen)
h = -(ave_shf + q + e) |
Measurements: | sgp15ebbrE13.b1: - Soil temperature 4 (resistance ratio)(soil_temp_4_rr)
more
sgp30ebbrE13.b1: - Surface soil heat flux 4(surface_soil_heat_flux_4)
- Surface sensible heat flux(sensible_heat_flux)
- Half hour averaged soil temperature 4(soil_temp_4)
- Surface soil heat flux, average of fluxes 1-5(surface_soil_heat_flux_avg)
- Change in energy storage 4, 0-5 cm soil layer(energy_storage_change_4)
- Surface latent heat flux(latent_heat_flux)
more
|