Description: | Testing of the batch of centrovision filters being used in the instrument indicate that
the first four filters leak 'out of band' light. The leakage for each filter is (1)415nm
3%, (2)500nm 1-2%, (3)615nm 1-2%, (4)673nm 3%. Acceptable leakage is less than 0.5%.
This leakage causes an error in the irradiance values (too high) at these four wavelengths
and the optical depths calculated using this data will be too high |
Measurements: | sgpnimfrC1.b1: - Normalized filter function data, filter 4, measured transmittance value obtained
during bench calibrations(normalized_transmittance_filter4) - Normalized filter function data, filter 3, measured transmittance value obtained
during bench calibrations(normalized_transmittance_filter3) - Normalized filter function data, filter 2, measured transmittance value obtained
during bench calibrations(normalized_transmittance_filter2) - Normalized filter function data, filter 1, measured transmittance value obtained
during bench calibrations(normalized_transmittance_filter1) - Direct horizontal narrowband irradiance, filter 1(direct_horizontal_narrowband_filter1)
- Narrowband Direct Horizontal Irradiance, Filter 2, cosine corrected(direct_horizontal_narrowband_filter2)
- Direct Horizontal Narrowband Irradiance, Filter 3(direct_horizontal_narrowband_filter3)
- Narrowband Direct Horizontal Irradiance, Filter 4, cosine corrected(direct_horizontal_narrowband_filter4)
- Narrowband Direct Normal Irradiance, Filter 1(direct_normal_narrowband_filter1)
- Narrowband Direct Normal Irradiance, Filter 3, cosine corrected(direct_normal_narrowband_filter3)
- Direct normal narrowband irradiance, filter 2, offset corrected(direct_normal_narrowband_filter2)
- Direct normal narrowband irradiance, filter 4, offset corrected(direct_normal_narrowband_filter4)
more
|