Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/19/1994 | 0000 | 11/27/1998 | 1930 |
Subject: | SGP/MWR/C1 - Reprocess: IRT insufficiently insulated |
DataStreams: | sgpmwrlosC1.b1, sgpmwrlosC1.a1 |
Description: | The downwelling IRT was insufficiently insulated to maintain an internal reference temperature above 0 degrees C. Measurements of sky temperature were over-estimated when instrument was below freezing. Data will be reprocessed, but users can correct data using the following correction factor: If T_reference < -5?C, then IRT_corrected = (IRT_original - 32.993K)/0.87238 where T_reference can be estimated with the ambient temperature (e.g. tkair) |
Suggestions: | |
Measurements: | sgpmwrlosC1.a1: more
sgpmwrlosC1.b1: more |