DQRID : D030311.1
Start DateStart TimeEnd DateEnd Time
03/03/2003050003/11/20030700
more
Subject:
TWP/SKYRAD/C2 - Reprocess: Incorrect UVB Calibration Coefficient
DataStreams:twpskyrad60sC2.b1, twpskyrad60sC2.a1, twpskyrad20sC2.a1
Description:
The uvb calibration coefficient was entered incorrectly as 0.2151
instead of 0.2332.
Suggestions: 
Multiply uvb output by 1.0841
Measurements:twpskyrad60sC2.a1:
  • Downwelling Ultra-Violet Irradiance, UV Biometer(down_uvb)
  • Downwelling Ultra-Violet Irradiance, UV Biometer, Maxima(down_uvb_max)
  • Downwelling Ultra-Violet Irradiance, UV Biometer, Standard Deviation(down_uvb_std)
  • Downwelling Ultra-Violet Irradiance, UV Biometer, Minima(down_uvb_min)