DQRID : D030427.2
Start DateStart TimeEnd DateEnd Time
04/08/2003234704/09/20030003
more
Subject:
TWP/SONDE/C1 - Questionable RH
DataStreams:twpsondewnpnC1.b1, twpsondewnpnC1.a1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 5km and 10km).  The RH oscillates rapidly between low (~5 %RH) and 
not so low (~50 %RH) and the oscillation seems to stop when the temperature is below ~35 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Suggestions: 
null
Measurements:twpsondewnpnC1.b1:
  • Dewpoint Temperature(dp)
  • Relative Humidity (hybrid-b grid)(rh)
more
twpsondewnpnC1.a1:
  • Dewpoint Temperature(dp)
  • Relative Humidity (hybrid-b grid)(rh)
more