Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2003 | 2347 | 04/09/2003 | 0003 |
Subject: | TWP/SONDE/C1 - Questionable RH |
DataStreams: | twpsondewnpnC1.b1, twpsondewnpnC1.a1 |
Description: | It appears that the RH sensor heating circuit failed during this portion of the sounding (approximately between 5km and 10km). The RH oscillates rapidly between low (~5 %RH) and not so low (~50 %RH) and the oscillation seems to stop when the temperature is below ~35 degC. According to Vaisala these are symptoms of failure in the ASIC (application specific integrated circuit) that controls the heating. |
Suggestions: | null |
Measurements: | twpsondewnpnC1.b1: more
twpsondewnpnC1.a1: more |