DQRID : D030515.3
Start DateStart TimeEnd DateEnd Time
05/13/2003190601/15/20050054
more
Subject:
NSA/MWR/C1  - no air temperature signal
DataStreams:nsamwrlosC1.b1, nsamwrC1.00, nsamwrtipC1.a1
Description:
When the new blower was upgraded by Radiometrics and reinstalled on the MWR, the air 
temperature sensor failed to properly report. It was determined that the wires carrying the 
signal to the analog board did not conform to the standard expected by the upgraded blower. 
The problem was corrected by changing the wiring and modifying the MWR software to read 
the signal from the appropriate corresponding channel.
Suggestions: 
Measurements:nsamwrtipC1.a1:
  • Ambient temperature(tkair)
more
nsamwrC1.00:
  • (Raw data stream - documentation not supported)
more
nsamwrlosC1.b1:
  • Ambient temperature(tkair)
more