Description: | Beginning on 06/03, the 415 nm channel appears to be erratic or noisy at times, which does
not coincide with the other channels. Maintenance on 07/02 corrected this problem.
head_temp readings occasionally spike or are erratic, especially between 1500 and 2000
UTC. On 06/17, head_temp readings became ambient due to a bad signal cable connection.
Maintenance on 07/02 ended this problem as well.
Due to the head_temp readings becoming largely out of tolerance, all channel readings
should be considered questionable, especially after 06/17. |
Measurements: | nsamfr10mC2.b1: - 10 meter Narrowband Upwelling Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
- 3 meter upwelling broadband hemispheric irradiance(up_hemisp_broadband)
- Detector temperature(head_temp)
- 25 meter upwelling narrowband hemispheric irradiance, filter 5, offset corrected(up_hemisp_narrowband_filter5)
- 3 meter upwelling narrowband hemispheric irradiance, filter 4(up_hemisp_narrowband_filter4)
- 3 meter upwelling narrowband hemispheric irradiance, filter 3, offset corrected(up_hemisp_narrowband_filter3)
- 3 meter upwelling narrowband hemispheric irradiance, filter 2, offset corrected(up_hemisp_narrowband_filter2)
- 10 meter upwelling narrowband hemispheric irradiance, filter 1(up_hemisp_narrowband_filter1)
more
nsamfr10mC2.00: - (Raw data stream - documentation not supported)
more
|