DQRID : D030730.3
Start DateStart TimeEnd DateEnd Time
12/16/2002170005/21/20031807
more
Subject:
SGP/IRT25m/C1 - error in data logger program
DataStreams:sgpirt25m20sC1.a0, sgpirt25mC1.a1
Description:
On 12/16/2002, the logger software for the IRT25m was
inappropriately edited. This resulted in incorrect instantaneous
values at 20 and 40 seconds after each minute and incorrect 
calculation of 60-second averaged instantaneous values (because
the 20 and 40 second values are two of three instantaneous
values used to calculate this average). On 5/21/2003, the data
logger program was corrected.
Suggestions: 
Only use the instantaneous values at 0 seconds after each minute
and ignore those at 20 and 40 seconds in sgpirt25m20sC1.
Do not use the 1-min averages of instantaneous data in sgpirt25mC1.a1.
Measurements:sgpirt25mC1.a1:
  • Instantaneous Case Temperature(inst_case_temp)
  • Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(inst_up_long_hemisp)
  • Instantaneous upwelling pyrgeometer case thermistor temperature, pyrgeometer(inst_up_long_case_temp)
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Instantaneous Dome Temperature(inst_up_long_dome_temp)
  • Instantaneous Surface Infra-red Temperature(inst_sfc_ir_temp)
  • Instantaneous Dome Temperature(inst_dome_temp)
more
sgpirt25m20sC1.a0:
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Surface Infra-red Temperature(inst_sfc_ir_temp)
  • Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyranometer Thermopile Voltage, Pyranometer(inst_up_short_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)
more