Subject:
SGP/MWR/C1 - no air temperature signal
DataStreams:sgpmwrlosC1.b1, sgpmwrtipC1.a1
Description:
When the new blower was upgraded by Radiometrics and reinstalled on the MWR, the air 
temperature sensor failed to properly report. It was determined that the wires carrying the 
signal to the analog board did not conform to the standard expected by the upgraded blower. 
The problem was corrected by changing the wiring.
Suggestions: 
Measurements:sgpmwrtipC1.a1:
  • Ambient temperature(tkair)
more
sgpmwrlosC1.b1:
  • Ambient temperature(tkair)
more