Subject: | SGP/MWR/B6 - thermal instability |
DataStreams: | sgpmwrlosB6.b1, sgpmwrlosB6.a1
|
Description: | During periods of high ambient temperature, the instrument may have been thermally
unstable. |
Suggestions: | The data are suspect when the blackbody temperature, tkbb, exceeds 319 K. |
Measurements: | sgpmwrlosB6.b1: - Mean 23.8 GHz sky brightness temperature(tbsky23)
- Mean 31.4 GHz sky brightness temperature(tbsky31)
- Total water vapor along LOS path(vap)
- Averaged total liquid water along LOS path(liq)
more
sgpmwrlosB6.a1: - Averaged total liquid water along LOS path(liq)
- Total water vapor along LOS path(vap)
- Mean 23.8 GHz sky brightness temperature(tbsky23)
- Mean 31.4 GHz sky brightness temperature(tbsky31)
more
|