Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/20/2004 | 2235 | 02/17/2004 | 2243 |
Subject: | SGP/MWR/B6 - Heater problem |
DataStreams: | sgpmwrtipB6.a1, sgpmwrlosB6.b1, sgpmwrlosB6.a1 |
Description: | When the heater/blower was replaced, the wet_window flag was constantly set high because the "moisture threshold voltage" value in the configuration file was not changed. The value was corrected for the new heater and the constant flagging stopped. |
Suggestions: | |
Measurements: | sgpmwrtipB6.a1: more
sgpmwrlosB6.b1: more sgpmwrlosB6.a1: more |