DQRID : D040722.2
Start DateStart TimeEnd DateEnd Time
11/01/2003000004/23/20041858
more
Subject:
NSA/METTWR/C1 - Chilled Mirror transmit logic board chipset unsuited for Arctic 
temperatures
DataStreams:nsamettwr4hC1.b1
Description:
Data missing due to problems with manufacturer's transmit logic board chipset.  They were 
generic chips not rated for Arctic temperatures.
Suggestions: 
Measurements:nsamettwr4hC1.b1:
  • Chilled Mirror Dew Point(CMHDP)
  • Chilled Mirror Calculated Saturation Vapor Pressure(SatVPCMH)
  • Chilled Mirror Temperature(CMHTemp)
  • Chilled Mirror Calculated Relative Humidity(CMHRH)
more