DQRID : D040904.2
Start DateStart TimeEnd DateEnd Time
05/24/2004000009/02/20042350
more
Subject:
TWP/SKYRAD/C3 - Incorrect NIP cal coefficient
DataStreams:twpskyrad60sC3.b1
Description:
During our site visit comparison at Darwin we observed that the skyrad NIP was reading 
about 50 w/m2 lower than the comparison cavity radfiometer and the replacement NIP on the 
comparison stand. We found that the calibration coefficient for the NIP (responsivity at 
45o)was put in as 8.64X10-6. The NIP was new and had not been calibrated at SGP. The 
factory number was 8.04X10-6. We corrected the calibration coefficient and the instruments 
agreed to better than 5 w/m2.
Suggestions: 
null
data for dates between 24 May 2004 and 2 Sept 2004 at Darwin by 1.072
Measurements:twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Shortwave direct normal irradiance, pyrheliometer, minima(short_direct_normal_min)
  • Shortwave direct normal irradiance, pyrheliometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
more