SGP/EBBR/E13 - Net Radiation Incorrect, Fluxes Incorrect
DataStreams:sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1
Description:
Rain entered the net radiometer through a crack in the upper dome and
settled in the bottom dome.  This caused too large net radiation measurements
during the following daylight hours, until the domes were replaced and dried out the same 
day.
Suggestions: 
Measurements:sgp15ebbrE13.b1:
  • Net radiation (mV)(net_radiation_mv)
more
sgp5ebbrE13.b1:
  • Net radiation(net_radiation)
more
sgp30ebbrE13.b1:
  • Surface sensible heat flux(sensible_heat_flux)
  • Surface latent heat flux(latent_heat_flux)
  • Net radiation(net_radiation)
more