Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/08/2004 | 1131 | 10/08/2004 | 1200 |
Subject: | TWP/SONDE/C1 - Bad RH at beginning of sounding |
DataStreams: | twpsondewnpnC1.b1 |
Description: | It appears that the RH sensor heating circuit failed during this portion of the sounding (approximately between 0km and 11km). The RH oscillates rapidly between low (~1 %RH) and high (~80 %RH). According to Vaisala these are symptoms of failure in the ASIC (application specific integrated circuit) that controls the heating. |
Suggestions: | null |