Subject: | SGP/NFOV/C1 - Condensation on window |
DataStreams: | sgpnfovC1.b1, sgpnfovC1.a0
|
Description: | The single-channel NFOV had a condensation problem. Water made its way into the interior
of the instrument and condensed on some of the important optical components of the system.
The condensation resulted in the signal strength of the 1-NFOV being significantly less
than that of the 2-NFOV. |
Suggestions: | Compare data with the 2-channel NFOV to determine on which days the problem occurred. |
Measurements: | sgpnfovC1.b1: - Calibrated zenith radiance data corrected for changes in head temperature.(radiance_temp_cor)
- Raw counts from instrument before calibration(raw_counts)
- Fields flagged by callang: 0 = No Flag
,(callang_flags)
- Uncalibrated Zenith Radiance(radiance)
more
sgpnfovC1.a0: - Uncalibrated Zenith Radiance(radiance)
- Ambient Temperature(ambient_temp)
- Fields flagged by callang: 0 = No Flag
,(callang_flags)
- Raw counts from instrument before calibration(raw_counts)
more
|