TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect
DataStreams:twpskyradC3.00, twpskyrad60sC3.b1
Description:
On 05/20 during installtion of the new Campbell datalogger system and associated program 
the NIP (S/N 33551E6) was assigned an incorrect calibration coefficient.  A coefficient of 
116.01 W/m2/mV was applied when it should have been 124.378 W/m2/mV.
TWP reports SV0405D and SV0409D detail the activities which lead to this DQR.
Suggestions: 
Back out original mV measurement and reapply correct cal coefficient.