DQRID : D050114.5
Start DateStart TimeEnd DateEnd Time
01/09/2005084901/18/20051900
more
Subject:
SGP/SIRS/E15 - Failed NIP cable
DataStreams:sgpsirsE15.00, sgpsirs20sE15.a0, sgpsirsE15.b1
Description:
After the ice storm, the SHORT_DIRECT_NORMAL values were bad or
              missing continuously. Corrective Maintenance on 01/18 found
              the NIP cable had broken inside the Bendix connector. The NIP 
              cable was repaired and returned to service.
Suggestions: 
Data is not recoverable. Disregard this data.
Measurements:sgpsirsE15.00:
  • (Raw data stream - documentation not supported)
more
sgpsirs20sE15.a0:
  • Instantaneous direct normal shortwave irradiance, pyrheliometer thermopile
    voltage(inst_direct_normal)
more
sgpsirsE15.b1:
  • Shortwave direct normal irradiance, pyrheliometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave direct normal irradiance, pyrheliometer, minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
more