Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/10/1997 | 2000 | 06/30/2003 | 2359 |
Subject: | SGP/EBBR/E18 - BA EBBR VAP Bulk Aerodynamics |
DataStreams: | sgp30baebbrE18.c1 |
Description: | Fixed AEM and wind speed sensor heights were improperly used in the calculation of bulk aerodynamic sensible and latent heat fluxes instead of site specific heights. This resulted in either under-estimation or over-estimation of the sensible and latent heat fluxes. |
Suggestions: | A rough adjustment to the fluxes is possible, but the preference is to correct the BA EBBR VAP program (to be addressed with an ECR), re-calculate the fluxes, and create revised data files in the ARM archive. |
Measurements: | sgp30baebbrE18.c1: more
|