Subject: | NSA/GNDRAD/C2 - Incorrect calibration coefficients applied to upwelling(Gndrad)PIR |
DataStreams: | nsagndradC2.00, nsagndrad60sC2.b1
|
Description: | An insitu test comparing the Gndrad PIR and the Gndrad IRT at the NSA was performed from 6
Jan 2005 to 12 Jan 2005. The radiometer mentor requested the Eppley single calibration
coefficient for the PIR replace the 4 coefficients method developed by NREL. The Grndrad
PIR compared closer to the Gndrad IRT during this period. However, the radiometer mentor
believes the 4 coefficient PIR irradiance measurements remains superior to the single
coefficient Eppley PIR measurement method. Further questions should be directed to the
radiometer instrument mentor. |
Suggestions: | Disregard this data. |
Measurements: | nsagndrad60sC2.b1: - Upwelling Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
- 10 meter Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
- PIR1 minimum(pir1_min)
- Upwelling longwave hemispheric irradiance, pyrgeometer, maxima(up_long_hemisp_max)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
Standard Deviation(up_long_hemisp_std) - PIR1 unshaded standard deviation(pir1_sd)
- PIR1 unshaded maximum(pir1_max)
- PIR1 mean(pir1_mean)
more
nsagndradC2.00: - (Raw data stream - documentation not supported)
more
|