Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2005 | 0258 | 04/01/2005 | 1545 |
Subject: | SGP/SMOS/E13 - T/RH sensor failure |
DataStreams: | sgp30smosE13.b1, sgp1440smosE13.b1, sgp1smosE13.b1 |
Description: | The T/RH sensor began to fail at high RH% values. Since this site has an aspirated radiation shield, values in excess of the 104% manufacturer limit are not expected to occur due to rain or condensation. |
Suggestions: | Use Co-located EBBR data or THWAPS data for RH% values. |
Measurements: | sgp1440smosE13.b1: more