Subject:
SGP/EBBR/E20 - Temp. of thum_left Incorrect At Times
DataStreams:sgp30ebbrE20.b1, sgp5ebbrE20.b1, sgp15ebbrE20.b1
Description:
The temperature portion of the left T/RH probe temperature spiked at times.  This was 
caused by a faulty surge protector in the J-Panel.