Subject: | SGP/NFOV/C1 - Head temp heater circuit failed |
DataStreams: | sgpnfovC1.b1, sgpnfovC1.a0
|
Description: | The head temp values were erratic. The instrument was removed from service and found the
head temp heater circuit failed in the full-on position frying the detectors. Data are
not useful. |
Suggestions: | Do not use data during this time period. |
Measurements: | sgpnfovC1.a0: - Detector temperature(head_temp)
- Ambient Temperature(ambient_temp)
- Tube Temperature(tube_temp)
- Fields flagged by callang: 0 = No Flag
,(callang_flags)
- Uncalibrated Zenith Radiance(radiance)
- Raw counts from instrument before calibration(raw_counts)
more
sgpnfovC1.b1: - Fields flagged by callang: 0 = No Flag
,(callang_flags)
- Tube Temperature(tube_temp)
- Cosine solar zenith angle(cosine_solar_zenith_angle)
- Detector temperature(head_temp)
- Uncalibrated Zenith Radiance(radiance)
- Calibrated zenith radiance data corrected for changes in head temperature.(radiance_temp_cor)
- Raw counts from instrument before calibration(raw_counts)
- Ambient Temperature(ambient_temp)
more
|