DQRID : D050414.1
Start DateStart TimeEnd DateEnd Time
04/12/2005000005/06/20051900
more
Subject:
SGP/NFOV/C1 - Head temp heater circuit failed
DataStreams:sgpnfovC1.b1, sgpnfovC1.a0
Description:
The head temp values were erratic.  The instrument was removed from service and found the 
head temp heater circuit failed in the full-on position frying the detectors.  Data are 
not useful.
Suggestions: 
Do not use data during this time period.
Measurements:sgpnfovC1.a0:
  • Detector temperature(head_temp)
  • Ambient Temperature(ambient_temp)
  • Tube Temperature(tube_temp)
  • Fields flagged by callang: 0 = No Flag ,(callang_flags)
  • Uncalibrated Zenith Radiance(radiance)
  • Raw counts from instrument before calibration(raw_counts)
more
sgpnfovC1.b1:
  • Fields flagged by callang: 0 = No Flag ,(callang_flags)
  • Tube Temperature(tube_temp)
  • Cosine solar zenith angle(cosine_solar_zenith_angle)
  • Detector temperature(head_temp)
  • Uncalibrated Zenith Radiance(radiance)
  • Calibrated zenith radiance data corrected for changes in head temperature.(radiance_temp_cor)
  • Raw counts from instrument before calibration(raw_counts)
  • Ambient Temperature(ambient_temp)
more

Close this window