Subject:
SGP/SIRS/E13 - Adjust Direct Normal Irradiance
DataStreams:sgpsirsE13.01, sgpsirsE13.00, sgp1swfanalsirs1longE13.c1, sgp15swfanalsirs1longE13.c1
Description:
Direct normal broadband shortwave irradiance measured by NIP s/n 30585E6 was found to have 
a 1.76% negative bias when compared in-situ with absolute cavity radiometer measurements 
taken on August 21, 2000.  The estimated measurement uncertainty of the calibration of 
NIP s/n 30585E6 on July 18, 2000 was -2.2% to +2.8%.  

Data from NIP s/n 30585E6 can be adjusted according to the in-situ comparisons to reduce 
the apparent measurement bias.
Suggestions: 
Multiply all direct normal irradiance measurements by 1.0176.