Description: | DATA REPROCESSED and CORRECTED:
During periods of high ambient temperature, the instrument was thermally unstable. The
problem was corrected by increasing the RF deck temperature from 50 to 56 C. The data were
reprocessed with derived calibration coefficients after the reference temperature tkmx was
adjusted.
ORIGINAL DQR CATEGORY: Yellow( Suspect)
REPROCESSED and CORRECTED DATA ARCHIVED:2016-03-29
REPROCESS IDENTIFICATION NUMBER: RID 206
---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------
During periods of high ambient temperature, the instrument may have been thermally
unstable. The problem was corrected by increasing the RF deck temperature from 50 to 56 C.
The data will need to be reprocessed with calibration coefficients derived after the
reference temperature, tkmx, was adjusted. |