DQRID : D050823.1
Start DateStart TimeEnd DateEnd Time
06/30/2005052806/30/20050711
more
Subject:
PYE/SONDE/M1 - Bad temperature and RH
DataStreams:pyesondewnpnM1.b1
Description:
The temperature appears to be approximately 10 degC too high during the entire sounding.  
This also causes the RH to be too low.  Most likely the problem was a bad temperature 
sensor.
Suggestions: 
No fix.
Measurements:pyesondewnpnM1.b1:
  • Relative Humidity (hybrid-b grid)(rh)
  • Dry Bulb Temperature(tdry)
  • Dewpoint Temperature(dp)
more