SGP/MWR/C1 - New software version (4.15) installed
DataStreams:sgpmwrlosC1.b1, sgpmwrtipC1.a1
Description:
A problem began with the installation of MWR.EXE version 4.12 in July 2002. The software 
had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier 
problem.  The software upgrade corrected the earlier problem but introduced a new one that 
caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of tip 
curves, and saturation of CPU usage. Software versions 4.13 and 4.14 also produced these 
problems.

The new MWR software, version 4.15, was installed on 08/04/2005. As a consequence of this 
upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to 
~50s.
Suggestions: 
Measurements:sgpmwrtipC1.a1:
  • Correlation coefficients for tip curve at 31.4 GHz(r31)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
  • 23.8 GHz Blackbody signal(bb23)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • Mixer kinetic (physical) temperature(tkxc)
  • Ambient temperature(tkair)
  • 23.8 GHz sky signal(tipsky23)
  • 23.8 GHz sky brightness temperature derived from tip curve(tbsky23tip)
  • (tknd)
  • 31.4 GHz sky brightness temperature derived from tip curve(tbsky31tip)
  • 23.8 GHz goodness-of-fit coefficient(r23)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • 31.4 GHz blackbody(bb31)
  • Blackbody kinetic temperature(tkbb)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • 31.4 GHz sky signal(tipsky31)
more
sgpmwrlosC1.b1:
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • Ambient temperature(tkair)
  • 23.8 GHz sky signal(sky23)
  • Total water vapor along LOS path(vap)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • Blackbody kinetic temperature(tkbb)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Averaged total liquid water along LOS path(liq)
  • 31.4 GHz blackbody(bb31)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Mean 31.4 GHz sky brightness temperature(tbsky31)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • 23.8 GHz Blackbody signal(bb23)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • Mixer kinetic (physical) temperature(tkxc)
  • (tknd)
  • Mean 23.8 GHz sky brightness temperature(tbsky23)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • 31.4 GHz sky signal(sky31)
  • Sky Infra-Red Temperature(sky_ir_temp)
more