SGP/EBBR/E8 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect
DataStreams:sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1
Description:
The top dome of the net radiometer was cracked and water had gotten inside it, so net 
radiation, and sensible and latent heat fluxes were incorrect.
Suggestions: 
Measurements:sgp5ebbrE8.b1:
  • Net radiation(net_radiation)
more
sgp15ebbrE8.b1:
  • Net radiation (mV)(net_radiation_mv)
more
sgp30ebbrE8.b1:
  • Surface latent heat flux(latent_heat_flux)
  • Net radiation(net_radiation)
  • Surface sensible heat flux(sensible_heat_flux)
more